P511 – Mirror Unit M3LB ELISA BESSY HZB
Application
Mirror unit (M3) for soft X-ray synchrotron radiation
Year of delivery
2024
Installation site
ELISA beamline (Low Branch) at BESSY II, Helmholtz-Zentrum Berlin, Germany
Design Features
-
Constant-strut-length hexapod mechanism.
- Six degrees of freedom and selectable rotation center.
- Spectral range: 50 – 3000 eV.
- Optical elements: one focusing mirror.
- Electrode at mirror surface for carbon contamination prevention.
- Temperature sensor at internal mechanics and at the mirror.
-
Four motorized and encoded OFHC blades at the chamber entrance flange; fluorescence coating of all blades.
- Mechanically decoupled internal mechanics from the vacuum chamber.
- Internal mechanics directly connected to the adjustable support structure underneath.
- Adjustable support structure.
- Permanently installed bakeout provisions.
Performance Features
- High step resolution.
- High pitch rotation stability and repeatability.
- Minimized mounting induced deformations.
Outer Dimensions
Technical specifications and performance values
Chamber type
cylindrical, vertical axis
Main flanges
at the top
Material
1.4404 (C < 0.02%)
Base pressure
10-10 mbar range
Beam height
1420 mm
Number of mirrors
1
Surface shape
cylindrical
Mirror dimensions
250 mm x 60 mm x 40 mm
Resolution (design value)
0.00157 µm /motor fine step
(1/128, recommended)
0.005 µm /encoder count
Range
± 8 mrad
Resolution (calculated value)
0.103 µrad
Resolution (measured value)
0.05 µrad
Repeatability
< 0.50 µrad
Range
± 10 mrad
Resolution (calculated value)
1.55 µrad
Resolution (meaasured value)
0.05 µrad
Repeatability
< 0.50 µrad
Range
± 10 mrad
Resolution (calculated value)
0.115 µrad
Range
± 20 mm
Resolution (calculated value)
5 nm
Range
± 20 mm
Resolution (calculated value)
39.7 nm /step
Resolution (measured value)
1 µm
Repeatability upon translation
0.50 µm
Range
± 0.50 mm
Resolution (calculated value)
34 nm
Material
OFHC
Translation range / blade
30 mm
Resolution (design value)
0.00157 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count
Resolution (measured value)
0.50 µm
Repeatability
< 1 µm