P477 - Mirror unit for VUV radiation based on constant-strut-length hexapod mechanism at NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany
P477 - Mirror unit for VUV radiation based on constant-strut-length hexapod mechanism at NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany

P477 – Mirror Unit NIM MLS PTB

Application

Mirror unit for VUV radiation

Year of delivery

2020

Installation site

NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany

Design Features

  • Constant-strut-length hexapod mechanism.

  • Six degrees of freedom and selectable rotation center.
  • Spectral range: 3 – 30 eV.
  • Optical elements: one refocusing mirror.
  • Mechanically decoupled internal mechanics from the vacuum chamber.
  • Internal mechanics directly connected to the adjustable support structure underneath.
  • EtherCAT based motion control system, synchronized encoder readout (suitable for on-the-fly measurements), software with server-client architecture, TCP-IP protocol, EPICS interface.

  • Adjustable support structure.
  • Permanently installed bakeout provisions.

Performance Features

  • High positioning resolution (minimum incremental motion).
  • High stability and repeatability for all degrees of freedom.
  • Minimized mounting induced deformations.

Outer Dimensions

P477 - Outer dimensions - Mirror unit for VUV radiation based on constant-strut-length hexapod mechanism at NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany

Technical specifications and performance values

General

Vacuum chamber

Chamber type

cylindrical, vertical axis

Main flanges

at the top

Material

1.4404 (C < 0.02%)

Base pressure

10-10 mbar range

Optical design parameters

Beam height

1630.30 mm

Number of mirrors

1

Surface shape

ellipsoidal

Mirror dimensions

600 x 120 x 60 mm3

Actuator drives for hexapod axes

Resolution (design value)

0.004 µm /motor fine step
(1/128, recommended)
0.005 µrad /encoder count

Mirror mechanics

Pitch rotation

Range

± 2°

Resolution (measured value)

1 µrad

Repeatability

0.16 µrad (pitch)

0.25 µrad (yaw)

Long-term stability (1h)

0.076 µrad

Roll rotation

Range

± 2°

Resolution (calculated value)

< 1 µrad

Yaw rotation

Range

± 2°

Resolution (measured value)

1 µrad

Repeatability

0.28 µrad (pitch)

0.18 µrad (yaw)

Long-term stability (1h)

0.053 µrad

Vertical translation

Range

± 5 mm

Resolution (calculated value)

< 0.10 µm

Horizontal translations

Range

± 5 mm

Resolution (calculated value)

< 0.10 µm

Performance Test Results

Pitch deviation after pitch change
P477 - Pitch deviation after pitch change - Mirror unit for VUV radiation based on constant-strut-length hexapod mechanism at NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany
Yaw deviation after pitch change
P477 - Yaw deviation after pitch change - Mirror unit for VUV radiation based on constant-strut-length hexapod mechanism at NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany
Pitch deviation after yaw change
P477 - Pitch deviation after yaw change - Mirror unit for VUV radiation based on constant-strut-length hexapod mechanism at NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany
Yaw deviation after yaw change
P477 - Yaw deviation after yaw change - Mirror unit for VUV radiation based on constant-strut-length hexapod mechanism at NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany
Mirror pitch positioning resolution
Mirror pitch stability (autocollimator measurement)
P477 - Mirror pitch stability - Mirror unit for VUV radiation based on constant-strut-length hexapod mechanism at NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany
Mirror yaw positioning resolution
Mirror yaw stability (autocollimator measurement)
P477 - Mirror yaw stability - Mirror unit for VUV radiation based on constant-strut-length hexapod mechanism at NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany
Mirror yaw positioning resolution
(auxiliary encoder, 1 nm resolution)
P477 - Mirror yaw positioning (auxiliary incremental encoder) - Mirror unit for VUV radiation based on constant-strut-length hexapod mechanism at NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany
Residual gas analysis
P477 - Residual gas analysis - Mirror unit for VUV radiation based on constant-strut-length hexapod mechanism at NIM beamline at Metrology Light Source, National Metrology Institute (PTB), Germany