P480 – PGM NSRL USTC

Application

Plane grating monochromator (PGM) for soft X-ray synchrotron radiation

Year of delivery

2020

Installation site

National Synchrotron Radiation Laboratory, University of Science and Technology, China

Design Features

  • Plane grating monochromator (PGM).
  • Spectral range: 200 – 1500 eV.
  • Optical elements: one plane mirror and two plane gratings.
  • Temperature sensors at internal mechanics, mirror and gratings holders.
  • Active water cooling of all optical elements.

  • Water cooled zero-order stop.
  • Motion control system with EPICS interface, EtherCAT, fast encoder readout suitable for on-the-fly measurements.

  • Mechanically decoupled internal mechanics from the vacuum chamber.
  • Internal mechanics directly connected to the adjustable support structure underneath.
  • Adjustable support structure.
  • Permanently installed bakeout provisions.

Performance Features

  • High positioning resolution (minimum incremental motion).
  • High pitch rotation stability and repeatability.
  • Minimized mounting induced deformations.

Outer Dimensions

P480 - Outer dimensions - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH

Technical specifications and performance values

General

Vacuum chamber

Chamber type

cylindrical, vertical axis

Main flanges

at the top

Material

1.4404 (C < 0.02%)

Base pressure

10-10 mbar range

Optical design parameters

Beam height (entrance)

1400 mm

Beam offset

20 mm

Number of plane mirrors

1

Plane mirror dimensions

400 mm x 40 mm x 50 mm

Number of gratings

2

Gratings dimensions

200 mm x 40 mm x 40 mm

Support structure

Adjustment ranges

Vertical translation

± 5 mm

Lateral translation

± 10 mm

Longitudinal translation

± 10 mm

Pitch

± 1°

Roll

± 1°

Yaw

± 1°

Plane mirror

Pitch rotation

Range

+0.6° to -16.52°

Resolution (design value)

0.0043 µrad /motor fine step
(1/128, recommended)
0.0097087 µrad /encoder count

Resolution (measured value)

0.10 µrad

Repeatability

0.22 µrad (pitch)
0.13 µrad (roll)

Long-term stability (> 4 h)

0.018 µrad

Translation

Range

-21 mm to +21 mm

Resolution (design value)

0.026 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Repeatability

0.32 µrad (pitch)
0.10 µrad (roll)

Gratings

Pitch rotation

Range

+0.6° to -16.92°

Resolution (design value)

0.0043 µrad /motor fine step
(1/128, recommended)
0.0097087 µrad /encoder count

Resolution (measured value)

0.10 µrad

Repeatability

0.13 µrad (pitch)
0.10 µrad (roll)

Long-term stability (> 7 h)

0.010 µrad

Scanning speed

1.12 mrad /s

Translation (grating exchange)

Range

-68.11 mm to +5.94

Resolution (design value)

0.026 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Repeatability

0.16 µrad (pitch)
0.40 µrad (roll)

Performance Test Results

Pitch deviation after pitch change (mirror)
P480 - Pitch deviation after pitch variation (mirror) - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Roll deviation after pitch change (mirror)
P480 - Roll deviation after pitch variation (mirror) - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Pitch deviation after pitch change (grating)
P480 - Pitch deviation after pitch variation (grating) - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Roll deviation after pitch change (grating)
P480 - Roll deviation after pitch variation (grating) - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Pitch deviation after mirror translation
P480 - Pitch deviation after mirror translation - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Roll deviation after mirror translation
P480 - Roll deviation after mirror translation - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Pitch deviation after grating translation
P480 - Pitch deviation after grating translation - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Roll deviation after grating translation
P480 - Roll deviation after grating translation - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Mirror pitch long-term stability
P480 - Mirror pitch stability - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Grating pitch long-term stability
P480 - Grating pitch stability - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Mirror pitch positioning resolution
P480 - Mirror pitch positioning - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Grating pitch positioning resolution
P480 - Grating pitch positioning - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Scanning speed of grating pitch
P480 - Scanning speed of grating pitch - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH
Residual gas analysis
P480 - Residual gas analysis - Plane grating monochromator (PGM) for soft X-ray synchrotron radiation at National Synchrotron Radiation Laboratory, University of Science and Technology, China - Bestec GmbH